Tuttnauer WTL198-0159

Tuttnauer WTL198-0159

Pre-sterilization TestingBowie & Dick and Helix Test Kits

Pre-sterilization test kits for monitoring air-removal and steam penetration.

KH2X15-3.5BD/M / Helix-PCD + Bowie-Dick simulation test kit for Steam sterilization processes: 3,5 min.134° C. Sterilizer monitoring. Metal capsule. Plastic tube of 1,5 m and 2 mm internal diameter. 1 cotton bag + 250 strips PCD-A-3.5BD Turns from Blue to Black. EN 867-5, ISO 11140-1 Type2. ISO 11140-4.
Designed to assess air removal and steam penetration in large and small vacuum-assisted steam sterilizers. Condition: Bowie & Dick cycle: 3.5 minutes, 134 ºC.
Characteristics
Helix-PCD
Metal capsule of great resistance and durability, to hold a Chemical Indicator designed for this system. The capsule is connected to a high resistance plastic tube of 1.5 m length (see catalogue for other lengths) and 2 mm internal diameter. Special silicone connector coated by external connector.
PCD-A-3.5BD
Type 2 Chemical Indicator.
Self-adhesive strip printed with indicator ink. Size: 76 x 6 mm.
10 strips per sheet. Size: 80 x 98.5 mm.
Initial color: blue.
Final color: dark grey/black.
Authorization: ANMAT (Argentinean National Administration of Drugs, Food and Medical Devices) PM-1614-4.
NOTE: If 4 bars change to the final color, the result is successful, the sterilizer works properly. If one or more bars have not turned to the final color, this shows a failure in sterilizing agent penetration, therefore an equipment malfunction

0.00

Description

Pre-sterilization TestingBowie & Dick and Helix Test Kits

Pre-sterilization test kits for monitoring air-removal and steam penetration.

KH2X15-3.5BD/M / Helix-PCD + Bowie-Dick simulation test kit for Steam sterilization processes: 3,5 min.134° C. Sterilizer monitoring. Metal capsule. Plastic tube of 1,5 m and 2 mm internal diameter. 1 cotton bag + 250 strips PCD-A-3.5BD Turns from Blue to Black. EN 867-5, ISO 11140-1 Type2. ISO 11140-4.
Designed to assess air removal and steam penetration in large and small vacuum-assisted steam sterilizers. Condition: Bowie & Dick cycle: 3.5 minutes, 134 ºC.
Characteristics
Helix-PCD
Metal capsule of great resistance and durability, to hold a Chemical Indicator designed for this system. The capsule is connected to a high resistance plastic tube of 1.5 m length (see catalogue for other lengths) and 2 mm internal diameter. Special silicone connector coated by external connector.
PCD-A-3.5BD
Type 2 Chemical Indicator.
Self-adhesive strip printed with indicator ink. Size: 76 x 6 mm.
10 strips per sheet. Size: 80 x 98.5 mm.
Initial color: blue.
Final color: dark grey/black.
Authorization: ANMAT (Argentinean National Administration of Drugs, Food and Medical Devices) PM-1614-4.
NOTE: If 4 bars change to the final color, the result is successful, the sterilizer works properly. If one or more bars have not turned to the final color, this shows a failure in sterilizing agent penetration, therefore an equipment malfunction

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.