Mindray A9 Anesthesia Machine

Mindray A9 Anesthesia Machine

Mindray A9 Anesthesia Machine

Physical Specifications
Dimensions and Weight
Height 1490 mm
Width 910 mm
Depth 705 mm
Weight 160 kg (with AG module and 3 yokes, without
vaporizers and gas cylinders)
Work Surface
Height 850 mm
Width 590 mm
Depth 325 mm
Weight limit 30 kg
Drawer (3 drawers, Internal Dimension)
Height 140 mm
Width 420 mm
Depth 315 mm
Weight limit 5 kg
Bag Arm
Height 1130 mm
Length 312 mm
Swiveling angle 120 degrees
Casters
Diameter 12.5 cm
Brake Centre brake system with Lock / Unlock icons
Cable pusher cable pusher wish each caster
Side mounting GCX Rails
Upper left length 130 mm
Upper right length 180 mm
Lower right length 485 mm
Supporting weight 27 kg at a maximum distance of 0.41 m
Work Light
Settings OFF, Low, High
Main Screen
Display size 18.5 inch
Display type Color LCD with capacitive touch screen
Resolution 1920 x 1080
Rotated 360 degrees
Tilted 60 degrees
Display parameters All setting and alarm parameters (including
Breath rate, I/E ratio, Tidal volume, Minute
volume, PEEP, MEAN, PEAK, PLAT, and O2
concentration, EtCO2, N2O, Aesthesia gas
concentration, BIS)
Graphic waveforms Pressure, Flow, Volume, CO2, O2, Anesthetic
gas, N2O, BIS, Pes, Ptp
Up to 5 waveforms display simultaneously
Spirometry loops Pressure-Volume, Flow-Volume and PressureFlow
Timer Display on screen timer
System status display
Display size 8.4 inch
Display type Color LCD
Resolution 800 x 600
Display content Volume exchanger indicator, gas supply
pressure, vaporizer status, AGSS status
Ventilator Specifications
Modes of Ventilation
Manual/Spontaneous ventilation/Bypass
Volume Control Ventilation (VCV) with PLV function
Pressure Control Ventilation (PCV)
Pressure Control Ventilation with volume guarantee (PCV-VG)
Continuous Positive Airway Pressure/Pressure Support Ventilation with
apnea backup (CPAP/PS)
Pressure Support Ventilation (PS) with apnea backup
Synchronized Intermittent Mandatory Ventilation
(SIMV-Volume Controlled and SIMV-Pressure Controlled)
Synchronized Intermittent Mandatory Ventilation Volume Guarantee
(SIMV-VG)
Airway Pressure Release Ventilation (APRV)
Adaptive Minute Ventilation (AMV)
Compensation
Circuit gas leakage compensation and automatic compliance
compensation
Ventilation Parameters Range
Patient type Adult, Pediatric, Neonate
Tidal volume 10 to 1500 mL ( VCV, SIMV-VC)
5 to 1500 mL (PCV-VG, SIMV-VG)
With TV/IBW indicator
Pinsp 5 to 90 cmH2O
Plimit 5 to 100 cmH2O
∆Psupp 0, 3 to 60 cmH2O (CPAP/PS)
Respiration rate 4 to 100 bpm
I:E 4:1 to 1:8
Tpause OFF, 5% to 60%
Tinsp 0.2 to 10.0 s
Trigger window 5% to 90%
Flow trigger 0.2 to 15 L/min
Pressure trigger -20 to -1 cmH2O
Exp% 5% to 80%
Min rate 2 to 60 bpm
Tslope 0.0 to 2.0 s
Apnea I: E 4:1 to 1:8
∆Papnea 3 to 60 cmH2O
Phigh 3 to 90 cmH2O
Plow 3 to 50 cmH2O
Thigh 0.2 to 10.0 s
Tlow 0.2 to 10.0 s
MV% 25% to 350%
Positive End Expiratory Pressure (PEEP)
Type Integrated, electronic controlled
Range 0 to 50 cmH2O
Monitoring Parameters
Tidal volume 0 to 3000 ml
Minute volume 0 to 100 L/min
Minute volume leakage 0 to 10.0 L/min
Peak pressure -20 to 120 cmH2O
Mean pressure -20 to 120 cmH2O
Plateau pressure -20 to 120 cmH2O
I:E 4:1 to 1:10
Rate 0 to 150 bpm
PEEP 0 to 70 cmH2O
Resistance (R) 0 to 600 cmH2O/(L/s)
Compliance (C) 0 to 300 ml/cmH2O
Inspired oxygen (FiO2) 18% to 100%
Control Accuracy
Volume delivery ≤60 ml: ± 10 ml
>60 ml and ≤ 210 ml: ±15 ml
>210 ml: ±7 % of the set value
Pressure delivery ± 2.0 cmH2O or ± 7% of the set value,
whichever is greater
PEEP ± 2.0 cmH2O or ± 7% of the set value,
whichever is greater
MV% ± 10% or ± 10% of the set value, whichever is
greater
Monitoring Accuracy
Volume monitoring ≤60 mL: ± 10 mL
>60 and ≤210 mL: ± 15 mL
>210 mL: ± 7% of the reading
Pressure monitoring ± 2.0 cmH2O or ± 4% of the reading,
whichever is greater
Rate ± 1bpm or ± 5% of the reading, whichever is
smaller
MV ± 0.1L/min or ± 8% of the reading, whichever
is greater
Alarm Setting
Paw High 2 to 100 cmH2O
Paw Low 0 to (Paw High –2) cmH2O
TV High 5 to 1600 mL
TV Low OFF, 0 to (TV High – 5) mL
MV High 0.2 to 100 L/min
MV Low 0 to 15 L/min: 0 to (MV High-0.2) L/min
15 to 100 L/min: 15 to (MV High – 1) L/min
FiO2 High 20% to 100%, OFF
FiO2 Low 18 to (FiO2 High – 2) %
Apnea alarm No breath has been detected within the
apnea time.
Apnea delay time 5 to 60 s (by volume or pressure)
10 to 40 s (by CO2 waveform)
Lung Recruitment Tool
Multi-step recruitment (Increasing PEEP progressively)
Control parameters a maximum of 7 steps
∆psupp, PEEP, Breaths, I:E, Rate
PEEP on exit
Preset procedure up to 5
One-step recruitment (sustain inflation)
Control parameters Pressure Hold, Hold Time, PEEP on exit
Cycle Interval OFF, 1 – 180 min
Auxiliary Pressure Monitoring
Monitor waveform Ptp, Pes
Monitor parameter PtpI, PtpE, ∆Ptp, PesI, PesE, ∆Pes
Data Storage and Recording
Configuration storage up to 10 customized profiles
Log storage 10000 entries of alarm and activity logs
History trend 48 hours of continuous trend data
Screenshot up to 50
Pre-use system check
Fully automatic performed by system, including hardware, flowmeter,
gas supply, power supply, module, breathing circuit leakage and
compliance, vaporizer and AGSS
Pneumatic Specifications
Pipeline Supply
Gas type O2, N2O and Air
Pipeline input range 280 to 600 kPa (40 to 87 psi)
Pipeline connections DISS or NIST
Pipeline Supply Pressure Monitoring
Display type Electronic
Ranges 0 to 1000kPa (0 to 140 psi)
Accuracy ± (4% of the full scale reading + 8% of the
actual reading)
Cylinder Supply
Cylinder supply E Cylinder (American style or UK style)
O2 input range 6.9 to 20 MPa (1000 to 2900 psi)
N2O input range 4.2 to 6 MPa (600 to 870 psi)
Air input range 6.9 to 20 MPa (1000 to 2900 psi)
Cylinder connections Pin-Index Safety System (PISS)
Yoke configuration O2, N2O, Air
Cylinder Supply Pressure Gauges
Display type Mechanical or Electronic
Air range 0 to 25 MPa (0 to 3500 psi)
O2 range 0 to 25 MPa (0 to 3500 psi)
N2O range 0 to 10 MPa (0 to 1400 psi)
Accuracy ± (4% of the full scale reading+8% of the
actual reading)
Ventilator Performance
Peak gas flow 180 L/min + Fresh Gas Flow
ACGO (Auxiliary Common Gas Outlet)
Control type Mechanical
Safety pressure A relief valve limits fresh gas pressure at ACGO
outlet port to not more than 12.5 kPa
O2 Flush
Flow rate 35 to 50 L/min
Auxiliary O2 & Air Flowmeter
Flow range 0 to 15 L/min
Oxygen concentration 21 % to 100 %
Indicator Glass tube and LED display
High Flow Nasal Cannula
Flow range 2 to 100 L/min
Oxygen concentration 21 % to100 %
Indicator Glass tube and LED display
Auxiliary High Pressure O2 Outlet
Pressure range 280 to 600 kPa
Maximum flow ≥ 90 L/min
O2 Controls
Supply failure alarm ≤ 220 kPa
Anesthetic Gas Scavenging System (AGSS)
Type of disposal system Passive
Active: High-flow or low-flow
Pump rate 75 to 105 L/min (High-flow)
25 to 50 L/min (Low-flow)
Management Scavenging flow rate monitoring and alarm
Automatically switch off when standby
Venturi Suction Regulator
Supply Air, from system gas source
Maximum vacuum ≥72 kPa at supply gas pressure of 280 kPa
≥73 kPa at supply gas pressure of 600 kPa
Maximum flow ≥25 L/min with pipeline gas at 280 kPa
≥32 L/min with pipeline drive gas at 600 kPa
Continuous Suction Regulator
Supply External vacuum
Maximum vacuum 517.5 mmHg to 540 mmHg (69 kPa to 72 kPa)
with external vacuum applied of 540 mmHg
and 40 L/min free flow
Maximum flow 39 L/min to 40 L/min with external vacuum
applied of 540mmHg and 40 L/min free flow
Electronic Flow control system (Electronic Mixer)
Direct Flow Control Mode
O2 flow range 0, 0.2 to 15 L/min
Air flow range 0 to 15 L/min
N2O flow range 0 to 12 L/min
O2 flow accuracy ± 50 ml/min or ± 5% of setting value,
whichever is greater
Balance gas (Air/N2O) flow accuracy
± 50 ml/min or ±5% of setting value,
whichever is greater
Total Flow Control Mode
Total flow range 0, 0.2 to 20 L/min
Total flow accuracy ± 100 ml/min or ± 5% of setting value,
whichever is greater
O2 concentration
Range 21% to 100% (The balance gas is Air)
26% to 100% (The balance gas is N2O)
Accuracy ± 5% V/V for flows < 1 L/min
± 5% of setting for flows ≥ 1 L/min
Optimizer
Available when AG module is loaded
Flow Pause
The fresh gas flow and ventilation will be paused for 1 minute at
default. (Maximum 2 minutes)
Backup Flow Control System
Control Type
Mechanical (Control needle valve and knob)
Flow Range
Control range (O2) 1 to 15 L/min
Control range (Air) 0 to 15 L/min
Total flow meter
Range 0 to 15 L/min
Indicator Flow tube
Indicator accuracy ± 10% of the indicated value for flows
(between 10% and 100% of full scale with
oxygen)
Breathing System Specification
Breathing system volume
Automatic ventilation 1800 ml
Manual ventilation 1950 ml
CO2 Absorber Assembly
Absorber capacity 1500 ml
Absorber type 1 Pre-Pak canister or
Loose Fill absorbent
Inspiratory Airway Pressure Gauge
Range -20 to 100 cmH2O
Accuracy ± (2% of the full scale reading + 4% of the
actual reading)
Flow Sensor
Type Variable orifice flow sensor
Location Inspiratory and expiratory port
Oxygen Sensor
Type Galvanic fuel cell
FiO2 displayed 18% to 100%
Accuracy ± (volume fraction of 2.5 % +2.5 % gas level)
Response time ≤20 seconds
Breathing System Connectors
Exhalation 22 mm OD / 15 mm ID conical
Inhalation 22 mm OD /15 mm ID conical
Manual bag port 22 mm OD /15 mm ID conical
Bag-to-Ventilator Switch
Type Bi-stable
Control Switch between manual and mechanical
ventilation
Adjustable Pressure Limiting (APL) Valve
Type Manually control with quick relief function
and illumination
Range Approximately (SP), 5 to 70 cmH2O
Tactile knob indication ≥ 30 cmH2O
Breathing Circuit Parameters
System compliance ≤ 2 mL/cmH2O
Volume of gas lost due to internal compliance
Impedance in manual mode ≤ 6 cmH2O
Impedance in automatic ventilation mode ≤ 6 cmH2O
Leakage ≤ 50 mL @ 3 kPa
System safety pressure on patient circuit 110 ± 10 cmH2O
Breathing System Temperature Controller
Breathing system temperature maintained at least 31°C typical at 20°C
ambient temperature in normal condition
Materials
All materials in contact with exhaled patient gases are autoclavable up
to a maximum temperature of 134°C, except O2 sensor and mechanical
pressure gauge.
All materials in contact with patient gas are latex free.
Vaporizers
Anesthetic Agent Delivery
Vaporizer V90 Electronic injection anesthetic vaporizer
Support agents Isoflurane, Sevoflurane, Desflurane
Position 2 positions
Filling system Safety filling adaptor (Iso, Sev)
Saf-T-Fill (Des)
Isoflurane range 0% to 5%
Sevoflurane range 0% to 8%
Desflurane range 0% to 18%
Control accuracy ± 15% of the setting or ±5% of full scale,
which is greater
Dimension
Height 215 mm
Width 75 mm
Depth 185 mm
Weight 3.5 kg (empty)
Agent capacity 320 mL
Liquid level Optical and electronic monitoring
Monitor Modules
Anesthesia Gas (AG) Module
Conformity with standard
ISO 80601-2-55
Measurement mode Infrared absorption, sidestream
Monitor gases CO2 , O2 (Paramagnetic O2 module), N2O, and
any of the five anesthetic agents: DES, ISO,
ENF, SEV and HAL
Warm-up time <45 s (ISO accuracy mode)
<10min (full accuracy mode)
Sample rate Adu/Ped: 150, 180, 200 ml/min
Neo: 100, 110, 120 ml/min
Monitoring range CO2:
0% to 30% (0.0 to 30 kPa, 0.0 to 226 mmHg)
O2/N2O: 0% to 100%
HAL, ISO, ENF: 0% to 30.0%
SEV: 0% to 30.0%
DES: 0% to 30.0%
BIS/BISx4 Module
Conformity with standard
IEC 60601-2-26
BIS, BIS L/ BIS R 0 to 100
Sweep speed 6.25 mm/s,12.5 mm/s, 25 mm/s or 50 mm/s
Alarm limit BIS high: 2 to 100
BIS low: 0 to (BIS high -2)
Calculated parameters SQI/SQI L, SQI R; EMG/EMG L, EMG R; SR/SR L,
SR R; SEF/SEF L, SEF R; TP/TP L, TP R; BC/BC L,
BC R; sBIS L, sBIS R; sEMG L, sEMG R; ASYM
www.mindray.com
P/N:ENG-A9 Datasheet -210285X4P-20200102
©2019 Shenzhen Mindray Bio-Medical Electronics Co.,Ltd. All rights reserved.
NMT Module
Conformity with standard
IEC 60601-2-10
Stimulation output
Pulse width: 100, 200, or 300 μs; monophasic rectangle
pulse; Accuracy: ± 10 %
Stimulation current range: 0 to 60 mA in increments of 5 mA
Accuracy: ± 5 % or ± 2mA, whichever is
greater
Maximum skin resistance: 3 kΩ @ 60 mA, 5 kΩ @ 40 mA
Block recovery OFF, 1,2, 3, 4, 5 %, 10 %, 20 %, 30 %, 40 %.
50 %, 60 %, 70 %,80 %, 90 %, 100 %
TOF (Train Of Four) mode
TOF-Ratio (response percentage) : 5 % to 160 %
TOF-Count (number of responses) : 0 to 4
TOF-T1% (response to the first stimulus as percentage of the
reference value) : 0 % to 200 %
ST (Single Twitch) mode
ST-Ratio (response percentage) : 0 % to 200 %
DBS (Double-Burst Stimulation) 3.2/3.3 mode
DBS-Ratio (response percentage) : 5 % to 160 %
DBS-Count (number of responses): 0 to 2
PTC (Post-Tetanic Count) mode
PTC-Count (number of responses) : 0 to 20
Anesthesia Function
Automatic controlled anesthesia (ACA)
Control range EtISO: 0% to 5%
EtSEV: 0% to 8%
EtDES: 0% to 18%
FiO2: 25% to 100%
Flow: Min, 0.3 to 6.0 L/min
Control accuracy EtAA: ± 0.2 Vol.% or ± 5% of the setting,
whichever is larger
FiO2: ± 3 Vol.% or ± 5% of the setting,
whichever is larger
Response time EtAA: <100s
FiO2: <130s (increase);
< 230s (decrease)
Stable time EtAA: <200s
FiO2: <140s (increase);
< 260s (decrease)
Anesthetic Prediction
Patient type Height: 150 to 200 cm
Weight: 40 to 140 kg
Age: 18 to 90 years old
Anesthetic agents Desflurane , Isoflurane, Sevoflurane
Prediction trend and waveform
Dynamic short trend waveforms of FiAA,
EtAA, FiO2 and EtO2 in the last 10 min
and prediction trend waveforms of FiAA,
EtAA, FiO2 and EtO2 in the next 20 min.
Prediction deviation EtAA=0: less than volume fraction of 0.05 %
EtAA≠0: – 20 % to 30 % of the measured EtAA,
or – 5 % to 7.5 % of the vaporizer maximum
setting, whichever is greater
EtO2 : – 10 % to 15 % of the measured EtO2, or
volume fraction of – 5 % to 7.5 %, whichever is
greater
Agent Consumption Calculation
Usage speed range ISO: 0 mL/h to 250 mL/h
SEV: 0 mL/h to 450 mL/h
DES: 0 mL/h to 900 mL/h
Accuracy ± 2 mL/h, or ± 15% of the reading, whichever
is larger
Total usage range 0 to 3000 ml
Accuracy ± 2 mL, or ± 15% of the reading, whichever is
larger
Electrical Specifications
Main Electrical Power
Power input 220-240 Vac, 50/60 Hz, 10A max
100-240 Vac, 50/60 Hz, 10A max
100-120 Vac,9 50/60 Hz, 10A max
Power cord 5 m (length)
Battery Power
Battery type Lead-acid, 12 VDC, 32 Ah (2 batteries)
Run-time New battery: minimum 90 minutes under
typical operating conditions
Time to shut down from the first Lower Battery Alarm
5 minutes minimum
(new fully-charged battery)
Battery charge time <12 hours
Auxiliary Electrical Outlets
Number of outlets 4
Output current 3 A for each outlet, 5 A for total
Communication Port
Communication port RS-232 compatible serial interface (DB9)
Network port Two RJ-45 network ports
USB port Four USB ports
Video signal port One VGA port for inputting the VGA video
signal of the main to external display
Environmental Specifications
Operating
Temperature 10 to 40°C
Relative humidity 15% to 95% (noncondensing)
Barometric (Kpa) 70 to 106.7 kPaz
Storage
Temperature -20 to 60°C for main unit,
-20 to 50°C for O2 sensor
Relative humidity 10% to 95% (noncondensing)
Barometric 50 to 106.7 kPa
Resistance to Ingress of Fluids
Complies with the requirements of clause 11.6.3 in IEC 60601-1 and
also the requirements in IEC 60529 for protection against vertically
falling water drops equipment (IPX1)

0.00

Description

Introducing the new safety
Mindray’s innovative electronic platform on the A9 empowers clinicians to ensure the safety of patients
throughout peri-operative periods, from induction to recovery, to improve patient outcome
With deep insights into the clinical workflow of the operating room, the A9’s user interface has been
designed around clinicians for reduced workloads and maximum patient safety
The A9 provides flexible data integration to meet the demands of various clinical scenarios. Compatible
with Mindray patient monitors and third-party clinical information systems, the A9 helps significantly
enhance the workflow efficiency in the operating room
Physical Specifications
Dimensions and Weight
Height 1490 mm
Width 910 mm
Depth 705 mm
Weight 160 kg (with AG module and 3 yokes, without
vaporizers and gas cylinders)
Work Surface
Height 850 mm
Width 590 mm
Depth 325 mm
Weight limit 30 kg
Drawer (3 drawers, Internal Dimension)
Height 140 mm
Width 420 mm
Depth 315 mm
Weight limit 5 kg
Bag Arm
Height 1130 mm
Length 312 mm
Swiveling angle 120 degrees
Casters
Diameter 12.5 cm
Brake Centre brake system with Lock / Unlock icons
Cable pusher cable pusher wish each caster
Side mounting GCX Rails
Upper left length 130 mm
Upper right length 180 mm
Lower right length 485 mm
Supporting weight 27 kg at a maximum distance of 0.41 m
Work Light
Settings OFF, Low, High
Main Screen
Display size 18.5 inch
Display type Color LCD with capacitive touch screen
Resolution 1920 x 1080
Rotated 360 degrees
Tilted 60 degrees
Display parameters All setting and alarm parameters (including
Breath rate, I/E ratio, Tidal volume, Minute
volume, PEEP, MEAN, PEAK, PLAT, and O2
concentration, EtCO2, N2O, Aesthesia gas
concentration, BIS)
Graphic waveforms Pressure, Flow, Volume, CO2, O2, Anesthetic
gas, N2O, BIS, Pes, Ptp
Up to 5 waveforms display simultaneously
Spirometry loops Pressure-Volume, Flow-Volume and PressureFlow
Timer Display on screen timer
System status display
Display size 8.4 inch
Display type Color LCD
Resolution 800 x 600
Display content Volume exchanger indicator, gas supply
pressure, vaporizer status, AGSS status
Ventilator Specifications
Modes of Ventilation
Manual/Spontaneous ventilation/Bypass
Volume Control Ventilation (VCV) with PLV function
Pressure Control Ventilation (PCV)
Pressure Control Ventilation with volume guarantee (PCV-VG)
Continuous Positive Airway Pressure/Pressure Support Ventilation with
apnea backup (CPAP/PS)
Pressure Support Ventilation (PS) with apnea backup
Synchronized Intermittent Mandatory Ventilation
(SIMV-Volume Controlled and SIMV-Pressure Controlled)
Synchronized Intermittent Mandatory Ventilation Volume Guarantee
(SIMV-VG)
Airway Pressure Release Ventilation (APRV)
Adaptive Minute Ventilation (AMV)
Compensation
Circuit gas leakage compensation and automatic compliance
compensation
Ventilation Parameters Range
Patient type Adult, Pediatric, Neonate
Tidal volume 10 to 1500 mL ( VCV, SIMV-VC)
5 to 1500 mL (PCV-VG, SIMV-VG)
With TV/IBW indicator
Pinsp 5 to 90 cmH2O
Plimit 5 to 100 cmH2O
∆Psupp 0, 3 to 60 cmH2O (CPAP/PS)
Respiration rate 4 to 100 bpm
I:E 4:1 to 1:8
Tpause OFF, 5% to 60%
Tinsp 0.2 to 10.0 s
Trigger window 5% to 90%
Flow trigger 0.2 to 15 L/min
Pressure trigger -20 to -1 cmH2O
Exp% 5% to 80%
Min rate 2 to 60 bpm
Tslope 0.0 to 2.0 s
Apnea I: E 4:1 to 1:8
∆Papnea 3 to 60 cmH2O
Phigh 3 to 90 cmH2O
Plow 3 to 50 cmH2O
Thigh 0.2 to 10.0 s
Tlow 0.2 to 10.0 s
MV% 25% to 350%
Positive End Expiratory Pressure (PEEP)
Type Integrated, electronic controlled
Range 0 to 50 cmH2O
Monitoring Parameters
Tidal volume 0 to 3000 ml
Minute volume 0 to 100 L/min
Minute volume leakage 0 to 10.0 L/min
Peak pressure -20 to 120 cmH2O
Mean pressure -20 to 120 cmH2O
Plateau pressure -20 to 120 cmH2O
I:E 4:1 to 1:10
Rate 0 to 150 bpm
PEEP 0 to 70 cmH2O
Resistance (R) 0 to 600 cmH2O/(L/s)
Compliance (C) 0 to 300 ml/cmH2O
Inspired oxygen (FiO2) 18% to 100%
Control Accuracy
Volume delivery ≤60 ml: ± 10 ml
>60 ml and ≤ 210 ml: ±15 ml
>210 ml: ±7 % of the set value
Pressure delivery ± 2.0 cmH2O or ± 7% of the set value,
whichever is greater
PEEP ± 2.0 cmH2O or ± 7% of the set value,
whichever is greater
MV% ± 10% or ± 10% of the set value, whichever is
greater
Monitoring Accuracy
Volume monitoring ≤60 mL: ± 10 mL
>60 and ≤210 mL: ± 15 mL
>210 mL: ± 7% of the reading
Pressure monitoring ± 2.0 cmH2O or ± 4% of the reading,
whichever is greater
Rate ± 1bpm or ± 5% of the reading, whichever is
smaller
MV ± 0.1L/min or ± 8% of the reading, whichever
is greater
Alarm Setting
Paw High 2 to 100 cmH2O
Paw Low 0 to (Paw High –2) cmH2O
TV High 5 to 1600 mL
TV Low OFF, 0 to (TV High – 5) mL
MV High 0.2 to 100 L/min
MV Low 0 to 15 L/min: 0 to (MV High-0.2) L/min
15 to 100 L/min: 15 to (MV High – 1) L/min
FiO2 High 20% to 100%, OFF
FiO2 Low 18 to (FiO2 High – 2) %
Apnea alarm No breath has been detected within the
apnea time.
Apnea delay time 5 to 60 s (by volume or pressure)
10 to 40 s (by CO2 waveform)
Lung Recruitment Tool
Multi-step recruitment (Increasing PEEP progressively)
Control parameters a maximum of 7 steps
∆psupp, PEEP, Breaths, I:E, Rate
PEEP on exit
Preset procedure up to 5
One-step recruitment (sustain inflation)
Control parameters Pressure Hold, Hold Time, PEEP on exit
Cycle Interval OFF, 1 – 180 min
Auxiliary Pressure Monitoring
Monitor waveform Ptp, Pes
Monitor parameter PtpI, PtpE, ∆Ptp, PesI, PesE, ∆Pes
Data Storage and Recording
Configuration storage up to 10 customized profiles
Log storage 10000 entries of alarm and activity logs
History trend 48 hours of continuous trend data
Screenshot up to 50
Pre-use system check
Fully automatic performed by system, including hardware, flowmeter,
gas supply, power supply, module, breathing circuit leakage and
compliance, vaporizer and AGSS
Pneumatic Specifications
Pipeline Supply
Gas type O2, N2O and Air
Pipeline input range 280 to 600 kPa (40 to 87 psi)
Pipeline connections DISS or NIST
Pipeline Supply Pressure Monitoring
Display type Electronic
Ranges 0 to 1000kPa (0 to 140 psi)
Accuracy ± (4% of the full scale reading + 8% of the
actual reading)
Cylinder Supply
Cylinder supply E Cylinder (American style or UK style)
O2 input range 6.9 to 20 MPa (1000 to 2900 psi)
N2O input range 4.2 to 6 MPa (600 to 870 psi)
Air input range 6.9 to 20 MPa (1000 to 2900 psi)
Cylinder connections Pin-Index Safety System (PISS)
Yoke configuration O2, N2O, Air
Cylinder Supply Pressure Gauges
Display type Mechanical or Electronic
Air range 0 to 25 MPa (0 to 3500 psi)
O2 range 0 to 25 MPa (0 to 3500 psi)
N2O range 0 to 10 MPa (0 to 1400 psi)
Accuracy ± (4% of the full scale reading+8% of the
actual reading)
Ventilator Performance
Peak gas flow 180 L/min + Fresh Gas Flow
ACGO (Auxiliary Common Gas Outlet)
Control type Mechanical
Safety pressure A relief valve limits fresh gas pressure at ACGO
outlet port to not more than 12.5 kPa
O2 Flush
Flow rate 35 to 50 L/min
Auxiliary O2 & Air Flowmeter
Flow range 0 to 15 L/min
Oxygen concentration 21 % to 100 %
Indicator Glass tube and LED display
High Flow Nasal Cannula
Flow range 2 to 100 L/min
Oxygen concentration 21 % to100 %
Indicator Glass tube and LED display
Auxiliary High Pressure O2 Outlet
Pressure range 280 to 600 kPa
Maximum flow ≥ 90 L/min
O2 Controls
Supply failure alarm ≤ 220 kPa
Anesthetic Gas Scavenging System (AGSS)
Type of disposal system Passive
Active: High-flow or low-flow
Pump rate 75 to 105 L/min (High-flow)
25 to 50 L/min (Low-flow)
Management Scavenging flow rate monitoring and alarm
Automatically switch off when standby
Venturi Suction Regulator
Supply Air, from system gas source
Maximum vacuum ≥72 kPa at supply gas pressure of 280 kPa
≥73 kPa at supply gas pressure of 600 kPa
Maximum flow ≥25 L/min with pipeline gas at 280 kPa
≥32 L/min with pipeline drive gas at 600 kPa
Continuous Suction Regulator
Supply External vacuum
Maximum vacuum 517.5 mmHg to 540 mmHg (69 kPa to 72 kPa)
with external vacuum applied of 540 mmHg
and 40 L/min free flow
Maximum flow 39 L/min to 40 L/min with external vacuum
applied of 540mmHg and 40 L/min free flow
Electronic Flow control system (Electronic Mixer)
Direct Flow Control Mode
O2 flow range 0, 0.2 to 15 L/min
Air flow range 0 to 15 L/min
N2O flow range 0 to 12 L/min
O2 flow accuracy ± 50 ml/min or ± 5% of setting value,
whichever is greater
Balance gas (Air/N2O) flow accuracy
± 50 ml/min or ±5% of setting value,
whichever is greater
Total Flow Control Mode
Total flow range 0, 0.2 to 20 L/min
Total flow accuracy ± 100 ml/min or ± 5% of setting value,
whichever is greater
O2 concentration
Range 21% to 100% (The balance gas is Air)
26% to 100% (The balance gas is N2O)
Accuracy ± 5% V/V for flows < 1 L/min
± 5% of setting for flows ≥ 1 L/min
Optimizer
Available when AG module is loaded
Flow Pause
The fresh gas flow and ventilation will be paused for 1 minute at
default. (Maximum 2 minutes)
Backup Flow Control System
Control Type
Mechanical (Control needle valve and knob)
Flow Range
Control range (O2) 1 to 15 L/min
Control range (Air) 0 to 15 L/min
Total flow meter
Range 0 to 15 L/min
Indicator Flow tube
Indicator accuracy ± 10% of the indicated value for flows
(between 10% and 100% of full scale with
oxygen)
Breathing System Specification
Breathing system volume
Automatic ventilation 1800 ml
Manual ventilation 1950 ml
CO2 Absorber Assembly
Absorber capacity 1500 ml
Absorber type 1 Pre-Pak canister or
Loose Fill absorbent
Inspiratory Airway Pressure Gauge
Range -20 to 100 cmH2O
Accuracy ± (2% of the full scale reading + 4% of the
actual reading)
Flow Sensor
Type Variable orifice flow sensor
Location Inspiratory and expiratory port
Oxygen Sensor
Type Galvanic fuel cell
FiO2 displayed 18% to 100%
Accuracy ± (volume fraction of 2.5 % +2.5 % gas level)
Response time ≤20 seconds
Breathing System Connectors
Exhalation 22 mm OD / 15 mm ID conical
Inhalation 22 mm OD /15 mm ID conical
Manual bag port 22 mm OD /15 mm ID conical
Bag-to-Ventilator Switch
Type Bi-stable
Control Switch between manual and mechanical
ventilation
Adjustable Pressure Limiting (APL) Valve
Type Manually control with quick relief function
and illumination
Range Approximately (SP), 5 to 70 cmH2O
Tactile knob indication ≥ 30 cmH2O
Breathing Circuit Parameters
System compliance ≤ 2 mL/cmH2O
Volume of gas lost due to internal compliance
Impedance in manual mode ≤ 6 cmH2O
Impedance in automatic ventilation mode ≤ 6 cmH2O
Leakage ≤ 50 mL @ 3 kPa
System safety pressure on patient circuit 110 ± 10 cmH2O
Breathing System Temperature Controller
Breathing system temperature maintained at least 31°C typical at 20°C
ambient temperature in normal condition
Materials
All materials in contact with exhaled patient gases are autoclavable up
to a maximum temperature of 134°C, except O2 sensor and mechanical
pressure gauge.
All materials in contact with patient gas are latex free.
Vaporizers
Anesthetic Agent Delivery
Vaporizer V90 Electronic injection anesthetic vaporizer
Support agents Isoflurane, Sevoflurane, Desflurane
Position 2 positions
Filling system Safety filling adaptor (Iso, Sev)
Saf-T-Fill (Des)
Isoflurane range 0% to 5%
Sevoflurane range 0% to 8%
Desflurane range 0% to 18%
Control accuracy ± 15% of the setting or ±5% of full scale,
which is greater
Dimension
Height 215 mm
Width 75 mm
Depth 185 mm
Weight 3.5 kg (empty)
Agent capacity 320 mL
Liquid level Optical and electronic monitoring
Monitor Modules
Anesthesia Gas (AG) Module
Conformity with standard
ISO 80601-2-55
Measurement mode Infrared absorption, sidestream
Monitor gases CO2 , O2 (Paramagnetic O2 module), N2O, and
any of the five anesthetic agents: DES, ISO,
ENF, SEV and HAL
Warm-up time <45 s (ISO accuracy mode)
<10min (full accuracy mode)
Sample rate Adu/Ped: 150, 180, 200 ml/min
Neo: 100, 110, 120 ml/min
Monitoring range CO2:
0% to 30% (0.0 to 30 kPa, 0.0 to 226 mmHg)
O2/N2O: 0% to 100%
HAL, ISO, ENF: 0% to 30.0%
SEV: 0% to 30.0%
DES: 0% to 30.0%
BIS/BISx4 Module
Conformity with standard
IEC 60601-2-26
BIS, BIS L/ BIS R 0 to 100
Sweep speed 6.25 mm/s,12.5 mm/s, 25 mm/s or 50 mm/s
Alarm limit BIS high: 2 to 100
BIS low: 0 to (BIS high -2)
Calculated parameters SQI/SQI L, SQI R; EMG/EMG L, EMG R; SR/SR L,
SR R; SEF/SEF L, SEF R; TP/TP L, TP R; BC/BC L,
BC R; sBIS L, sBIS R; sEMG L, sEMG R; ASYM
www.mindray.com
P/N:ENG-A9 Datasheet -210285X4P-20200102
©2019 Shenzhen Mindray Bio-Medical Electronics Co.,Ltd. All rights reserved.
NMT Module
Conformity with standard
IEC 60601-2-10
Stimulation output
Pulse width: 100, 200, or 300 μs; monophasic rectangle
pulse; Accuracy: ± 10 %
Stimulation current range: 0 to 60 mA in increments of 5 mA
Accuracy: ± 5 % or ± 2mA, whichever is
greater
Maximum skin resistance: 3 kΩ @ 60 mA, 5 kΩ @ 40 mA
Block recovery OFF, 1,2, 3, 4, 5 %, 10 %, 20 %, 30 %, 40 %.
50 %, 60 %, 70 %,80 %, 90 %, 100 %
TOF (Train Of Four) mode
TOF-Ratio (response percentage) : 5 % to 160 %
TOF-Count (number of responses) : 0 to 4
TOF-T1% (response to the first stimulus as percentage of the
reference value) : 0 % to 200 %
ST (Single Twitch) mode
ST-Ratio (response percentage) : 0 % to 200 %
DBS (Double-Burst Stimulation) 3.2/3.3 mode
DBS-Ratio (response percentage) : 5 % to 160 %
DBS-Count (number of responses): 0 to 2
PTC (Post-Tetanic Count) mode
PTC-Count (number of responses) : 0 to 20
Anesthesia Function
Automatic controlled anesthesia (ACA)
Control range EtISO: 0% to 5%
EtSEV: 0% to 8%
EtDES: 0% to 18%
FiO2: 25% to 100%
Flow: Min, 0.3 to 6.0 L/min
Control accuracy EtAA: ± 0.2 Vol.% or ± 5% of the setting,
whichever is larger
FiO2: ± 3 Vol.% or ± 5% of the setting,
whichever is larger
Response time EtAA: <100s
FiO2: <130s (increase);
< 230s (decrease)
Stable time EtAA: <200s
FiO2: <140s (increase);
< 260s (decrease)
Anesthetic Prediction
Patient type Height: 150 to 200 cm
Weight: 40 to 140 kg
Age: 18 to 90 years old
Anesthetic agents Desflurane , Isoflurane, Sevoflurane
Prediction trend and waveform
Dynamic short trend waveforms of FiAA,
EtAA, FiO2 and EtO2 in the last 10 min
and prediction trend waveforms of FiAA,
EtAA, FiO2 and EtO2 in the next 20 min.
Prediction deviation EtAA=0: less than volume fraction of 0.05 %
EtAA≠0: – 20 % to 30 % of the measured EtAA,
or – 5 % to 7.5 % of the vaporizer maximum
setting, whichever is greater
EtO2 : – 10 % to 15 % of the measured EtO2, or
volume fraction of – 5 % to 7.5 %, whichever is
greater
Agent Consumption Calculation
Usage speed range ISO: 0 mL/h to 250 mL/h
SEV: 0 mL/h to 450 mL/h
DES: 0 mL/h to 900 mL/h
Accuracy ± 2 mL/h, or ± 15% of the reading, whichever
is larger
Total usage range 0 to 3000 ml
Accuracy ± 2 mL, or ± 15% of the reading, whichever is
larger
Electrical Specifications
Main Electrical Power
Power input 220-240 Vac, 50/60 Hz, 10A max
100-240 Vac, 50/60 Hz, 10A max
100-120 Vac,9 50/60 Hz, 10A max
Power cord 5 m (length)
Battery Power
Battery type Lead-acid, 12 VDC, 32 Ah (2 batteries)
Run-time New battery: minimum 90 minutes under
typical operating conditions
Time to shut down from the first Lower Battery Alarm
5 minutes minimum
(new fully-charged battery)
Battery charge time <12 hours
Auxiliary Electrical Outlets
Number of outlets 4
Output current 3 A for each outlet, 5 A for total
Communication Port
Communication port RS-232 compatible serial interface (DB9)
Network port Two RJ-45 network ports
USB port Four USB ports
Video signal port One VGA port for inputting the VGA video
signal of the main to external display
Environmental Specifications
Operating
Temperature 10 to 40°C
Relative humidity 15% to 95% (noncondensing)
Barometric (Kpa) 70 to 106.7 kPaz
Storage
Temperature -20 to 60°C for main unit,
-20 to 50°C for O2 sensor
Relative humidity 10% to 95% (noncondensing)
Barometric 50 to 106.7 kPa
Resistance to Ingress of Fluids
Complies with the requirements of clause 11.6.3 in IEC 60601-1 and
also the requirements in IEC 60529 for protection against vertically
falling water drops equipment (IPX1)